To enable accurate measurements inside blood components, PLService developed measurement phantoms based on precise temperature loggers with a probe located inside the container, together with proprietary reporting software.
We have successfully used these phantoms for plasma thawing validation for several years. Users appreciate their carefully considered design and measurement precision. In validation of blood component temperatures, the key parameter is the actual temperature inside the product, rather than the surrounding temperature of thawing, refrigeration or freezing equipment.
We have frequently encountered blood bank staff struggling to meet annual validation duties. In 99% of such cases, this did not result from a lack of knowledge or skills, but from a lack of suitable tools and solutions for conducting the process in accordance with guidelines and obtaining reliable, repeatable results.
PLService engineers offer a complete solution. Contact us.
What distinguishes PLService measurement phantoms?
- Temperature measurement inside the product.
- Validation does not interfere with everyday equipment use.
- Measurement reflects the actual effects of opening and closing equipment, transport containers and rooms on the stored product’s temperature.
- A digital report with a temperature curve, protected against third-party alteration of recorded data.
- Compared with conventional sensors placed inside equipment, measurement fluctuations caused by compressor switching are eliminated.
- Assessment of additional parameters such as safe power-off time, daily hysteresis and the effects of external temperature.
- Up to 2,730 recorded points per process at 1- or 5-minute intervals, depending on the procedure.
- Phantom volume close to the product volume: 250 ml ±25 ml, except for platelet concentrates.
- Measurement range: −40°C to +80°C.
- Measurement accuracy: ±0.25°C.
- Measurement resolution: 0.03°C.
Services using measurement phantoms
Validation for serology laboratories and blood banks